7

How to use

STEP 1. Gently place the cup of the
device over the ear. It's important
that the edges of the cup are in
contact with the patient's head.
Move any hair out of the way,
remove glasses, earrings and
hair bands or anything that may
prevent achieving adequate
contact around the ear.
STEP 2. Press the start button
once to initiate the functional test
sequence. If it is working correctly,
you will see all 3 lights flash in
sequence 3 times. This indicates
that the device is ready to use.
The test sequence will begin
in 3 seconds with three signals
at 1000 Hz. (55dBHL, 35dBHL,
20dBHL).
A light will appear when a tone
is being played: first left, then
middle, then right. The subject
should indicate when a tone is
heard to the tester by a suitable
method such as raising a finger.