Siemens HearCheck Screener User Manual page 10

Table of Contents

Advertisement

7
STEP 3. Count and record the number of times indicated for
the first sequence.
STEP 4. The start button must be pushed within 20 seconds
of the end of the first sequence (after the third light has
appeared) to start the second sequence.
STEP 5. You will see all 3 lights flash in sequence 3x3 times.
The test sequence will begin in 3 seconds with three
signals at 3000 Hz. (75dBHL, 55dBHL, 35dBHL).
STEP 6. Count and record the number of times the
subject has indicated that a tone was heard.
STEP 7. Repeat steps 2 through 6 on the other ear.
10