Care and cleaning

The probe tip is the most delicate part of the thermometer.
It must be clean and intact to ensure accurate readings.
If the thermometer is ever accidentally used without a lens filter,
clean the probe tip as follows:
Very gently wipe the surface with a cotton swab or soft cloth moistened with
alcohol. After the alcohol has completely dried out, you can put a new lens
filter on and take a temperature measurement.
If the probe tip is damaged, please contact your local service centre.
Use a soft, dry cloth to clean the thermometer display and exterior. Do not use
abrasive cleaners. Never submerge this thermometer in water or any other liquid.
Store thermometer and lens filters in a dry location free from dust and contamination
and away from direct sunlight.
Additional lens filters (LF40) are available at most stores carrying Braun ThermoScan

Replacing the batteries

Insert new batteries when the battery symbol appears
1
on the display.
Open the battery compartment. Remove the batteries and replace
2
with new batteries, making sure the poles are in the right direction.
3
Snap battery door into place.
Only discard empty batteries. To protect the environment, dispose of empty
batteries at your retail store or at appropriate collection sites according to
national or local regulations.
Calibration
The thermometer is initially calibrated at the time of manufacture. If this thermometer
is used according to the use instructions, periodic re-adjustment is not required. If at
any time you question the accuracy of the temperature measurement, please contact
your authorized service center.
Manufacturing date is given by the LOT number located in the battery compartment.
The first three (3) digits after LOT represent the Julian date that the product was
manufactured and the next two (2) digits represent the last two numbers of the
calendar year the product was manufactured. The last identifiers are the letters that
represent the manufacturer.
An example: LOT 11614K, this product has been manufactured on the 116th day of the
year 2014.
8

Errors and troubleshooting

Error Message
.
®
Situation
Solution
No lens filter is attached.
Attach new, clean lens filter.
The thermometer probe was not
Make sure that the positioning of the
positioned securely in the ear.
probe is correct and remains stable.
An accurate measurement
was not possible.
Replace lens filter and reposition.
Press Start button to begin a new
POS = position error
measurement.
Ambient temperature is not within
Allow the thermometer to remain for
the allowed operating range
30 minutes in a room where
(10 – 40 °C or 50 – 104 °F).
the temperature is between 10
and 40 °C or 50 and 104 °F.
Temperature taken is not within
Make sure the probe tip and lens are
typical human temperature range
clean and a new, clean lens filter is
(34 – 42.2 °C or 93.2 – 108 °F).
attached. Make sure the thermometer
is properly inserted. Then, take a new
HI = too high
temperature.
LO = too low
Wait 1 minute until the thermometer
System error – self-check display flashes
turns off automatically, then turn
continuously and is not followed by the
on again.
ready beep and the ready symbol.
... reset the thermometer by removing
If error persists,
the batteries and putting them back in.
... please contact your local service
If error still persists,
centre.
Battery is low, but thermometer will still
Insert new batteries.
operate correctly.
Battery is too low to take correct
Insert new batteries.
temperature measurement.