T5-600/T5-1000
Electrical Tester

Parts and Service

The tester is warranted to be free from defects in material and workmanship for two years, while under normal
use. Parts and repairs are warranted for 90 days. For the complete warranty statement, refer to the
T5-600/T5-1000 Electrical Tester Instruction Sheet.

Specifications

Accuracy is specified for one year after calibration, at 18 °C to 28 °C (64 °F to 82 °F) with relative humidity
to 90 %. AC conversions are ac-coupled, average responding, and calibrated to the rms value of a sine wave
input. Accuracy specifications are given as follows:
Temperature coefficient of 0.1 x (specified accuracy)/ °C for <18 °C or >28 °C (<64.4 °F or >82.4 °F)
Serial Numbers 79560000 and Above
General Specifications
Calibration
Maximum Voltage
Between any Terminal
and Earth Ground
Maximum Voltage
Between Current Fork
and Earth Ground
Temperature
Altitude
Relative Humidity
Battery Type and Life
Shock, Vibration
Surge Protection
Enclosure Rating
2
Shop for Fluke products online at:
±([ % of reading] + [number of least significant digits])
One-year calibration cycle.
T5-600: 600 V rms, Overvoltage Category III, Pollution Degree 2
T5-1000: 1000 V rms, Overvoltage Category III, 600 V RMS, OverVoltage Category IV,
Pollution Degree 2
T5-600 and T5-1000: 1000 V rms, Overvoltage Catergory III, also 600 V rms, Overvoltage
Category IV, Pollution Degree 2, (This dual rating applies to the current fork only.)
Operating: -10 °C to +50 °C (14 °F to 122 °F); Storage: -30 °C to +60 °C (-22 °F to +140 °F)
Operating: 2000 m (6562 ft); Storage: 10,000 m (32808 ft)
0 % to 95 %, 5 °C to 30 °C (41 °F to 86°F); 0 % to 75 %, 30 °C to 40 °C (86 °F to 104 °F);
0 % to 45 %, 40 °C to 50 °C (104 °F to 122 °F)
AA (2); 360 hours continuous with alkaline; 125 hours continuous with zinc chloride
1 m drop at -10 °C to + 50 °C (14 °F to 122 °F) per ANSI/ISA-S82.01-1994 and EN 61010-1
1995. Random vibration per MIL-PRF-28800F for a Class 2 instrument
(5 Hz to 55 Hz, 3 g maximum)
T5-600: 6 kV per IEC 61010, T5-1000: 8 kV per IEC 61010
IP 52 per IEC 60529, no vacuum applied