Endress+Hauser iTEMP HART TMT182 Operating Instructions Manual

Endress+Hauser iTEMP HART TMT182 Operating Instructions Manual

Temperature head transmitter
Hide thumbs Also See for iTEMP HART TMT182:

Advertisement

Quick Links

Operating instructions
®
®
iTEMP
HART
TMT182
Temperature head transmitter
6
BA105R/24/ae/04.05
51006279

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser iTEMP HART TMT182

  • Page 1 Operating instructions ® ® iTEMP HART TMT182 Temperature head transmitter BA105R/24/ae/04.05 51006279...
  • Page 2 TMT182 Endress+Hauser...
  • Page 3: Table Of Contents

    10.9 Human interface ......33 10.10 Certificates and approvals ....33 Endress+Hauser...
  • Page 4: Safety Message

    / process information in conflict with any patent. Please note that Endress+Hauser reserves the right to change and / or improve the product design and specifications without notice.
  • Page 5: Short Form Instructions

    Æ Trouble-shooting / fault-finding page 25 If problems occur after commissioning or during operation always start fault finding using the check list. Special questions will act as a guide to the cause of the fault and the necessary cure. Endress+Hauser...
  • Page 6: Safety Notes

    Third party validated metrics are available for the temperature transmitter. Testing is done per IEC 61508 for Safety Instrumented Systems. The safety manual can be separated ordered separatly under order code: SD 006R09EN More details and download see: www.us.endress.com Endress+Hauser...
  • Page 7: Operational Safety

    If one of these icons is on the device’s nameplate, the device can be used in hazardous areas. Hazardous area This symbol identifies the hazardous area in the diagrams in these Operating Instructions. – Devices that are used in hazardous areas or cables for such devices must have the corresponding type of protection. Endress+Hauser...
  • Page 8: Identification

    Measurement, Control, Regulation and Laboratory Procedures" and with the EMC requirements of IEC61326. The measuring system described in these Operating Instructions thus complies with the statutory requirements of the EC Directives. Endress+Hauser confirms successful testing of the device by affi- xing to it the CE mark. â...
  • Page 9: Registered Trademarks

    ® ® • iTEMP and ReadWin 2000 are registered trademarks of Endress+Hauser Wetzer GmbH + Co. KG, Nesselwang, Germany Delivery contents Delivery contents The delivery contents of a temperature head transmitter are as follows: • Head transmitter TMT182 • Installation screws, installation springs and E-ring •...
  • Page 10: Installation

    • Screw the head transmitter into the field housing using a screwdriver. " Caution! In order to avoid damaging the head transmitter, do not over-tighten the installation screws. ƒig. 4: Installation of transmitter into a sensor connection head (left side) and a field housing TAF10 (right side). Endress+Hauser...
  • Page 11: Typical North American Installation

    ƒig. 5: : Installing the head transmitter - Head transmitter assembly Pos. 1: Thermowell Pos. 2: Measuring inset Pos. 3: Extension neck Pos. 4: Transmitter housing Pos. 5: Head transmitter assembly Pos. 6: Installation screws Pos. 7: Head transmitter Endress+Hauser...
  • Page 12: Wiring

    Open the cable gland on the head transmitter or field housing. Feed the cable through the opening in the cable gland and then connect the cable wires to terminals 1 and 2 according to the wiring diagram (see fig. 6). Note! The screws on the terminals must be screwed in tightly. Endress+Hauser...
  • Page 13: Hart ® Connection

    = Commubox FXA191 (in combination with a PC - E+H operating software ’Commuwin II’), a* = Commubox FXA191 connected to the communication sockets of a power supply unit, b = Loop power supply (e.g. RNS221 power supply or ® RN221N active barrier), c = HART transmitter, d = PLC with a passive input Endress+Hauser...
  • Page 14: Potential Grounding

    (see page 18). Note! ® • When using the HART hand unit all parameters can be read out, however, programming is ® locked. It is possible to access the HART function matrix by entering 281 in the LOCK function. Endress+Hauser...
  • Page 15: Commuwin Ii-Operating Program

    • Data security of unit parameters • Unit diagnostics • Measurement point documentation Note! More detailed information on Commuwin II can be found in the E+H documentation: • Operating manual: BA124F/00/en “Commuwin II” operating program More details see: www.us.endress.com Endress+Hauser...
  • Page 16: Commissioning

    (see fig. 11) is possible. Using the Commuwin II operating matrix (see fig. 12) a quick setup is possible with the following fields: • Type of sensor (V2H0) • Unit meas. value (V2H2) • Value for 4 mA (V2H4) • Value for 20 mA (V2H5) • Connection (V2H6) Endress+Hauser...
  • Page 17: Configuration With Hart® Protocol And Dxr275, Dxr375 Hand Operated Module

    ® • More detailed information on the HART hand operated module can be found in the respective operating manual in the carrying case. ƒig. 10: : Configuration of hand module example, "Sensor input" Endress+Hauser...
  • Page 18: Commuwin Ii

    • Data security of unit parameters • Unit diagnostics • Measurement point documentation Note! More detailed information on Commuwin II can be found in the E+H documentation: • Operating manual: BA 124F/00/en “Commuwin II” operating program More details see: www.us.endress.com Endress+Hauser...
  • Page 19: Configuration Using Hart® Protocol

    – Output (4 to 20 mA/20 to 4 mA) – Damping (filter) – Offset – TAG (Measurement point description) – Identifier (Descriptor) Service functions – Simulation (on/off) – Reset/factory default – Series number (only display) – Operation code (=release code 281) Endress+Hauser...
  • Page 20: Description Of Unit Functions

    Entry of zero point correction (Offset). • V0H5 Entry: -18.00 to 18.00 °F (10.00 to 10.00 °C) (°C, °F, K) 0.00 °C Note! Entry returns to factory default values when changing sensor type! a. Only for Commuwin II operating matrix Endress+Hauser...
  • Page 21 • V2H6 Entry: 2 wire 3 wire • RTD connection 4 wire 3 wire Note! Function field is only active on selection of resistance thermometer (RTD) in the unit function SENSOR TYPE (V2H0) a. Only for Commuwin II operation matrix Endress+Hauser...
  • Page 22 Note! Min. process value will be changed to the actual process value on access. On reset to factory default, the default value is entered. +10000 a. Only for Commuwin II operation matrix b. Not for thermocouples (TC) Endress+Hauser...
  • Page 23 • VAH2 e.g.: 1.0000 indicates version 1.00.00. Software Version Display of software version, • VAH3 e.g.: 8010 indicates version 1.0. • Software Rev. Serial number 8-digit display of E+H device serial numbers (see legend plates on the device). • VAH4 Endress+Hauser...
  • Page 24: Supported Hart® Commands

    Enter/Exit fixed primary varaible current mode Perform master reset Write primary variable units Read additional transmitter status Write number of response preambles Write burst mode command number Burst mode control E+H specific Read matrix parameter Write matrix parameter Upload Download Endress+Hauser...
  • Page 25: Maintenance

    Fault code Cause Action/cure No fault, Warning None Hardware fault (unit defective) Replace head transmitter Sensor short circuit Check sensor Sensor cable open circuit Check sensor Reference measurement point defec- None tive Unit not calibrated Return head transmitter to manufacturer Endress+Hauser...
  • Page 26: Application Faults Without Messages

    Change parameter ‘Connection type’ number) Transmitter programming faulty (scale) Change scale Wrong RTD used Change parameter ‘Sensor type’ Sensor connection (2 wire) Check sensor connections Sensor cable (2 wire) not compensated Compensate cable resistance Offset incorrectly set Check offset Endress+Hauser...
  • Page 27: Spare Parts

    Please follow the return Authorization Policy which is attached with these instructions. Disposal Disposal Due to its construction, the head transmitter cannot be repaired. When disposing of the head trans- mitter please take note of the local disposal regulations. Endress+Hauser...
  • Page 28: Software History

    (DXR275, DXR375) or PC (COMMUWIN II or ReadWin 2000 operating software). 10.2 Input Measured variable Temperature (temperature linear transmission behavior), resistance and voltage Measuring range The transmitter monitors different measuring ranges depending on the sensor connection and input signals. Endress+Hauser...
  • Page 29: Output

    Signal (mA) Under ranging Standard Over ranging Standard 20.5 ≤ 3.6 Sensor break; sensor short circuit low To NAMUR NE 43 ≥ 21.0 Sensor break; sensor short circuit high To NAMUR NE 43 Endress+Hauser...
  • Page 30: Power Supply

    ≤ 3.6 mA is generated. After approx. 2 to 3 s, the system makes another attempt to output the signal corresponding to the temperature. ≤ 3 V at U ≥ 13 V, f Residual ripple Allowable ripple U = 1 kHz max. Endress+Hauser...
  • Page 31: Performance Characteristics

    Influence of cold junction Pt100 IEC 60751 Cl. B (internal cold junction for thermocouples TC) 10.6 Installation conditions Installation instructions • Installation angle: no limit • Installation area: Connection head accord. to DIN 43 729 Form B; TAF10 field housing Endress+Hauser...
  • Page 32: Environmental Conditions

    Line interference IEC 61000-4-16 10 kHz to 150 kHz 10 V 10.8 Mechanical construction Design, dimensions ƒig. 14: : Dimensions of the head transmitter in inches (mm) Weight approx. 1.4 oz (40 g) Material • Housing: PC • Potting: PUR Endress+Hauser...
  • Page 33: Human Interface

    • IEC 61010: Safety requirements for electrical measurement, control and laboratory instrumenta- tion. • IEC 61326: Electromagnetic compatibility (EMC requirements) • NAMUR: Standardization association for measurement and control in chemical and pharmaceu- tical industries. (www.namur.de) • NEMA: Standardization association for the electrical industry Endress+Hauser...
  • Page 34 Windows® 2000 ....... 16 Installation angle ....... . 9 Endress+Hauser...
  • Page 35 TMT182 Endress+Hauser...
  • Page 36 Canada México Instruments International Endress+Hauser, Inc. Endress+Hauser Canada Endress+Hauser, México, S.A. de C.V. Endress+Hauser 2350 Endress Place 1075 Sutton Drive Av. Gustavo Baz No. 43 Instruments International AG Greenwood, IN 46143 Burlington, ON L7L 5Z8 Fracc. Bosques de Echegaray Kaegenstrasse 2 Canada Naucalpan de Juárez, C.P.