Bosch vivatmo pro Instructions For Use Manual page 46

Hide thumbs Also See for vivatmo pro:
Table of Contents

Advertisement

Maintenance and troubleshooting
en
en
E‑1
en
E‑2
en
en
E‑4
E‑5
en
en
E‑6
en
E‑7
en
en
en
E‑8
en
46
Measurement procedure is incorrect,
breathing was too strong, too weak or time-
limit after regeneration was exceeded.
Permissible number of measurement trials
reached soon, measurements still possible.
Number of permissible measurement
trials reached. Device does not perform
any measurements after maximum
number of permissible tests is reached.
Humidity in device too high.
Ambient temperature and air pressure
(altitude) is outside the specified range.
Battery is too weak for measurement.
System self-test failed
Regeneration failed.
© Bosch Healthcare Solutions GmbH
Repeat the measurement, see "4.2 Measurement stand-alone
with the handheld".
See display for number of remaining measurement trials. Order
new Vivatmo pro handheld, see "7.2 Status information handheld".
Connect a new Vivatmo pro handheld and dispose the old one,
see chapter "8 Disposal".
Check ambient conditions and take a new mouthpiece.
Use the Vivatmo pro in location with permissible ambient
conditions.
When transporting the device from one place to another, a stabi-
lization period of at least 20 minutes should be observed before use.
Place the device on the basestation for charging.
Make sure the battery is sufficiently charged by positioning the
handheld on the basestation for at least 2 hours.
Take out and re-insert the battery of the handheld if the error
persits and/or change the battery against a new one.
If the error persists, contact your distribution partner.
Press the ENTER button to repeat the regeneration.
If the errors persists, contact your distribution partner.