Settings
QC test Level 0
en
en
Jeppe M. Gregersendan
Exp. Date 2018-09-10
25 PPB
en
QC-Measurements
2016-04-07
11 : 43
en
2015-10-07
13: 16
2015-06-21
09: 55
en
Level 0 Measurement
en
Last measurement : 2018-03-12 – 07:02
0 PPB
en
The measurement was successful.
en
C O N T I N U E
en
en
en
en
38
If Level 0 QC is configured perform the following steps:
P E R F O R M Q C
1. Select your user name and select the PERFORM QC button
(only if QC tester is activated).
2. Release the handheld device from the basestation and attach a Level 0
mouthpiece.
Passed
Passed
Passed
3. Perform the measurement, see "4.1 Measurement using the basestation".
When the result is 0 ppb, the Level 0 QC test was successful.
If the result is 5 ppb or higher the Level 0 QC test failed.
Repeat the measurement and take care to use a new Level 0 mouthpiece.
When the Level 0 QC test still fails, contact the service.