To qualify as tester, you must complete the following steps:
en
Perform 4 measurements within 7 days, not more than one qualification measurement per day.
The QC-measurement on the fourth day must be within accepted range of 5 ppb ±3 × standard deviation, at least ±3 ppb
en
from the mean value.
en
INFORMATION
en
en
Qualification procedure
en
QC Tester
en
Language
Jeppe M. Gregersendan
Administrator
en
Konsta Haapakoski
Professional
Oona Pakarinen-Kattilakoski
en
Professional
en
en
en
36
The moving mean value is recalculated when the QC tester performs a new QC-measurement within
7 days.
The qualification of a QC tester expires after 30 days. Then the qualification is suspended and the QC
tester needs to qualify again, according to the qualification procedure.
Note: In the QC settings, the QC Tester control must be activated,
see "6.4 Administrator settings". Perform the following steps:
1. In the Settings menu, select PERFORM QC.
Qualifiying
1/3 Passed
The QC Tester screen opens and the actual user is highlighted and can
Qualified
be selected.
2. Select the QUALIFY button.