Replace CrystalPress Sorter Bowl
For Kit 110688-90, Procedure P/N 110907 Rev 0
Tools required: Phillips screw driver, an 11/32"
socket wrench and a grounding wrist strap.
Step 1:
Turn off, unplug the power cord and
communication cable.
Step 3: From the sorter bowl that you need to re-
place, unscrew the 3 locking nuts (counter clockwise).
:
Step 5
Using the 11/32" socket wrench, unscrew
(counter clockwise), the nylon locking nut from the
front of the bowl as shown.
Step 7: Pull the sorter bowl towards you to clear the
threaded stud, and then lift slightly up, then pull the
bowl out and place it on the table as shown. (Wires
are still connected to the PCB board at this
point).
You will need to support and balance the Crys-
talPress when you turn it up-side down.
Step 2: Turn the CrystalPress over, remove the 7
screws, take the bottom pan off and place aside.
Step 4: All 3 locking nuts are removed. No need to
remove the silver screw (arrow).
You can turn the CrystalPress back on its legs.
Step 6: Sorter bowl is shown with the locking nut
removed. (You do not need to remove the 2
screws shown by the arrows)
Step 8: IMPORTANT: WRIST STRAP MUST BE
WORN UNTIL FINISHED WITH THE DISCON-
NECTIONS ON THE BOARD.
1
14140 NE 200th St.
Woodinville, WA 98072
1.425.398.8282
www.ioline.com