Agilent Technologies 11713B Operating And Service Manual page 81

Attenuator/switch drivers
Table of Contents

Advertisement

N O T E
11713B/C Operating and Service Manual
Step
Action
3 Verify LAN
a At IO SETUP screen,
status on
select IP ADDR
11713B/C
softkey
b At IP ADDRESS
screen, select LAN
STAT softkey
c Verify values
CURRENT IP
ADDRESS, CURRENT
SUBNET and
CURRENT GATEWAY
are valid
d Ensure LAN STATUS
is NORMAL
e If LAN STATUS is
FAULT, check your
LAN connection and
then press Restart
Netw softkey and wait
f Press BACK softkey to
go back to IP
ADDRESS screen
Now you can use the 11713B/C Web Browser Interface to access and
control the instrument. See
Over LAN"
on page 85.
If you plan to program the instrument over LAN or use such
programs, make sure you have installed the required I/O software on
your computer.
Use the Connectivity Expert utility of the Agilent IO Libraries Suite
to add the 11713B/C and verify a connection. When identifying the
instrument, it is easiest if you use the IP address noted in step 3
above.
If you have installed any other I/O software, refer to documentation included with
that software.
Now you can use various programming environments to control the
11713B/C. For an overview on programming instructions via USB,
refer to the Connectivity Guide.
Remote Interface Configurations
Illustration
"Exploring the 11713B/C Web Interface
7
81