Remote Operations
70
To Set Supply Voltage
The commands below set the supply voltage for each bank.
• CONFigure:BANK1 {OFF|P5v|P15v|P24v|USER}
• CONFigure:BANK2 {OFF|P5v|P15v|P24v|USER}
To Query Supply Voltage
The commands below query the supply voltage set for each bank.
Value to be returned: OFF, P5, P15, P24 or USER.
• CONFigure:BANK1?
• CONFigure:BANK2?
To Query Number of Relay Cycles
The command below query the number of relay cycles for each
channel. Value to be returned: number of cycles.
• DIAGnostic:RELay:CYCles? <channel list>
To Clear Relay Cycles
The commands below clear relay cycle for selected or all channels.
• DIAGnostic:RELay:CLEAr <channel list>
• DIAGnostic:RELay:CLEAr:ALL