How to Specify Events and Reactions to Events
HP 81200 Data Generator/Analyzer Platform User Guide, Revision 2.1
5 In the GOTO column, select one of the available blocks.
Figure 77
Branch Table – Block Selection
Note that the END block is an implicit block which is always available. It
terminates the sequence.
The branch table of block 2 in our example is shown below:
Figure 78
Branch Table – Block 2
As soon as CMD1 occurs, the block will not be repeated any more. The
sequence will execute the Idle block.
6 Decide whether you wish to generate a trigger pulse. If you wish to
activate the VXI triggers for controlling other VXI modules, ensure that
you have not defined an event based on the status of the VXI trigger
lines.
7 Click OK.
To complete the example, you have to repeat the steps 1 to 6 for block 3.
Figure 79
Branch Table – Block 3
As soon as CMD0 occurs, the infinite loop of the Idle block will be left, and
the sequence will execute the Measure block
Creating the Stream of Generated and Expected Data