6. Record the AM ME ER value on the test record.
7. Set the function generator for a 1 kHz, 1.800 V p-p, sine wave output
signal. If necessary, adjust the function generator for a DVM reading of
0.6363 AC V rms.
8. Record the AM ME ER value on the test record.
9. Disconnect the BNC tee from the AM input and connect it to the FM
input.
10. On the synthesizer, turn o external AM, turn on external FM, set the FM
sensitivity to 1 MHz/V, and turn on the FM meter function. Press:
AM Menu AM On/Off Ext
MOD
5
FM Menu FM On/Off Ext
PRIOR
5
1
MHz
5
5
Monitor Menu Meter On/Off FM
PRIOR
5
11. Set the function generator for a 1 kHz, 10 V p-p, sine wave output signal.
If necessary, adjust the function generator amplitude for a DVM reading of
3.535 AC V rms.
12. Record the FM ME ER value on the test record.
13. Set the function generator for a 100 kHz, 2 V p-p, sine wave output signal.
14. Record the FM ME ER value on the test record.
Related Adjustments
None
In Case of Difficulty
Refer to the
Troubleshooting
2-134 Performance Tests