May 2019
3.2
Gas flow applications
Procedure
1. Place taps in the top or side of the line.
2. Mount level or above the taps.
3.3
Steam flow applications
Procedure
1. Place taps to the side of the line.
2. Mount beside or below the taps.
3. Fill impulse lines with water.
3.4
Environmental seal for housing
Thread sealing (PTFE) tape or paste on male threads of conduit is required to
provide a water/dust tight conduit seal and meets requirements of NEMA
Type 4X, IP66, and IP68. Consult factory if other ingress protection ratings
are required.