3
Chapter 3 Features and Functions
Burst Mode
• When the gated mode is selected, the burst count, burst period,
and trigger source are ignored (these parameters are used for the
triggered burst mode only). If a manual trigger is received, it is
ignored and no error will be generated.
• When the gated mode is selected, you can also select the polarity of
the signal on the rear-panel Ext Trig connector.
• Front-Panel Operation: After enabling bursts, press the N Cycle
(triggered) or Gated softkey.
Remote Interface Operation:
[SOURce[1|2]:]BURSt:MODE {TRIGgered|GATed}
Use the following command to select the polarity of the external gate
signal on the Ext Trig connector. The default is NORM (true-high
logic).
[SOURce[1|2]:]BURSt:GATE:POLarity {NORMal|INVerted}