Cleaning; Disinfecting - Invacare HF2RE9ES User Manual

1.7l (m9) cylinder with continuous flow regulator for homefill oxygen system
Table of Contents

Advertisement

Invacare 1.7L (M9) Cylinder With Continuous Flow Regulator
5. Examine the cylinder gauge H until the pressure is
below 500 psig (3447 kPa).
Cylinder pressure is below 500 psig (3447 kPa)
when the cylinder gauge needle J is in the red
zone.
6. Turn the flow dial E to setting "0".
7. Close the shutoff valve G fully.
Turn the shutoff valve clockwise to close.

5.3 Cleaning

Refer to the safety information in the Maintenance
section before performing this procedure.
Follow the cleaning/disinfectant product manufacturer's
instructions for use, handling, storage and disposal of the
product.
Follow this procedure to clean the cylinder assembly as
required when visibly soiled.
1. Ensure the cover is installed on the fillport.
2. Clean exterior surfaces of the cylinder with a lint free
cloth dampened with 70% isopropyl alcohol.
3. Clean the exterior surfaces of the regulator with a dry,
lint free cloth.
4. If the regulator is still visibly soiled, clean with a lint
free cloth dampened with 70% isopropyl alcohol. DO
NOT wipe the fillport or outlet fitting.
5. Wipe the cylinder and regulator with a dry lint free
cloth. Ensure the product is dry before using.
22

5.4 Disinfecting

Refer to the safety information in Maintenance
before performing this procedure.
Follow the cleaning/disinfectant product manufacturer's
instructions for use, handling, storage and disposal of the
product.
This procedure is for disinfecting the cylinder assembly and
is to be performed by the provider prior to a patient change
and as otherwise required. Recommended disinfectant
products:
Quaternary ammonium based wipes selected by the
provider for the desired disinfecting characteristics
70% isopropyl alcohol wipes
1. Ensure the cover is on the fillport.
2. If heavily soiled, first clean the cylinder assembly. Refer
to the Cleaning section.
3. Use a wipe to disinfect the external surfaces of
the regulator and cylinder. Follow the disinfectant
manufacturer's instructions. DO NOT wipe the fillport
or outlet connection.
4. Wipe the cylinder and regulator with a dry lint free
cloth. Ensure the product is dry before using.
1167426-E