Invacare 1.7L (M9) Cylinder With Continuous Flow Regulator
1. Prior to each use, inspect product for visible damage.
Refer to the External Examination section. DO NOT use
if any damage is found.
If for any reason, any label becomes illegible or
lost, contact your dealer/Home Care equipment
provider.
2. Inspect the cylinder gauge H to ensure the pressure
is at least 1500 psig (10342 kPa). If below 1500 psig
(10342 kPa), then refill the cylinder assembly using the
HomeFill system.
WARNING!
– When needle on cylinder gauge enters
red zone, it is recommended to refill the
cylinder assembly on the Invacare HomeFill
Compressor.
3. Position the cylinder assembly A so the outlet
connection D is pointing away from the user and any
other person(s).
WARNING!
Risk of Injury
To reduce the risk of injury:
– DO NOT direct flow of oxygen at any
person, or flammable material when
adjusting the flow dial E.
4. Turn the flow dial E to setting "0".
5. Remove the outlet connection cap, if present.
18
6. Attach a standard or high flow single lumen adult
oxygen nasal cannula with a maximum length of 2.1 m
(7 feet) to the product's outlet connection according to
the cannula manufacturer's instructions.
DO NOT use pediatric cannulas with this product.
DANGER!
Risk of Injury or Death
The cannula can cause tripping, falling, or
other injury if improperly positioned and
secured. To avoid injury or death:
– The cannula MUST be routed and secured
properly.
– DO NOT position the cannula around the
neck. Ensure the patient can move freely
while wearing the cannula.
– Avoid positioning cannulas across areas
of high foot traffic (i.e. aisles, doorways,
hallways, etc.)
7. Place the cannula over your ears and position the
prongs in your nose as instructed by your health care
provider or cannula manufacturer.
1167426-E