Invacare HF2RE9ES User Manual page 11

1.7l (m9) cylinder with continuous flow regulator for homefill oxygen system
Table of Contents

Advertisement

DANGER!
Risk of Fire
Dirt or contaminants on the cylinder assembly or
regulator can cause fire, injury or death.
– DO NOT use cleaning/disinfecting solutions on
the fillport fitting.
– DO NOT immerse product in any kind of liquid.
– DO NOT allow any kind of liquid to get inside
openings of the regulator.
– DO NOT spray cleaning/disinfecting solutions
on product.
– DO NOT autoclave.
– DO NOT gas sterilize with ethylene oxide.
– DO NOT attempt to repair cylinder assembly.
All repairs MUST be performed by Invacare
Corporation or in an Invacare approved repair
center.
– Store cylinder assembly in a clean, dry and cool
area when not in use.
DANGER!
Risk of Injury or Death
To avoid choking or ingestion of chemicals from
airway contamination:
– DO NOT use the product in the presence
of pollutants, smoke, fumes, flammable
anesthetics, cleaning agents, or chemical
vapors.
1167426-E
WARNING!
Risk of Injury or Death
Contaminants in the oxygen may cause injury or
death.
– DO NOT use if the oxygen has an unusual odor.
– DO NOT use if particulates are inhaled.
– Contact Invacare or dealer for service.
WARNING!
Risk of Injury or Damage
To avoid injury or damage from product misuse:
– The flow dial should be turned with normal
pressure from one's hand. If the flow dial is
stiff to turn, do not force, otherwise damage
could occur.
– DO NOT use a flow outlet for driving any
medical equipment.
Safety
11