Omron E8F2 Instruction Manual page 12

Pressure sensor
Hide thumbs Also See for E8F2:
Table of Contents

Advertisement

E8F2
Auto-teaching
One-point Teaching (Window-mode Teaching)
OUT1
Entered
or
Not entered
1. Press the SET Key under condition 3 below while t.w is displayed to display the present measurement value.
One-point Teaching
Condition 3
ON
OFF
ON point
2. Check the measurement value and press the Set Key to perform and complete teaching.
3. The values are entered Press the SET Key for approximately 2 s while t.w is displayed to enter the values and return the E8F2 to
measurement mode.
4. Press the Up and Down Keys simultaneously to return the E8F2 to measurement mode without selecting the values.
Note: The E8F2 will be automatically set to window mode by performing one-point teaching.
One-point teaching is ideal for applications that check original pressure.
Teaching Error
Present value not
within set range
Meaning of display: er.t (Error teach)
If the present value or the result of calculation after teaching is not
within the range, the SET Key input will not be accepted, in which
case an error will be displayed for 1 s if teaching is performed.
12
Present value
Measurement mode
ON point: TEACH1 – 30% FS
OFF point: TEACH1 + 30% FS
The default value is 10% FS (changeable).
Pressure value
OFF point
Error display for 1 s
E8F2