The actual current in the conductor is 400 mA divided by 10-
turns, or 40 mA.

Maintenance

Performance Test

Verify Probe accuracy by measuring the output of a 20 A (±0.2
%), 60 Hz current source (Fluke 5100B and 5220A or
equivalent). When used with a compatible DMM with ± 0.75 % or
better ac current accuracy (Fluke 8060A or equivalent), the probe
should measure 20.00 A ± 0.40 A (20.00 mA, ± 0.40 mA). There
are no calibration adjustments in the Clamp.
Cleaning
Clean the case periodically by wiping it with a damp cloth and
detergent. Do not use abrasive cleaners or solvents. Do not
immerse the Clamp in liquids.

Specifications

Electrical Characteristics

Current Range
Continuous Measuring Range
Overload Capability
for 1 Hour, 33 % Duty Cycle
Output Sensitivity
Turns Ratio
Accuracy (at 100 A, 2 kHz, +25 °C) ± 1% of reading
Load Impedance
Phase Shift at 2 kHz
Frequency Range
800 A rms
100 mA to 800 A rms
1500 A rms (50/60 Hz)
1 mA / A
1000:1
≤ 20 Ω
± 1 degree
30 Hz to 10 kHz (-3 dB)