Heartbeat Coincidence

Enabling/Disabling Heartbeat Coincidence Detection

Display Indicator

Revision D
Fetal Heart Rate Monitoring: Heartbeat Coincidence
Heartbeat coincidence is available on Models 172, 173, and 174 (dual heart rate
monitors) to alert you when you may be monitoring a duplicate signal. Heartbeat
coincidence is indicated when two heartbeats have a consistent phase relationship
for equal to or greater than 60% of the detected beats for about 60 seconds; the
cessation of coincidence is indicated when the phase relationship is inconsistent for
greater than 40% of the detected beats for about seven seconds.
Table 5-1
summarizes the combinations of heart rate sources that are continuously
compared for the possibility of coincidence.
Table 5-1. Heartbeat Coincidence Comparisons
Mode
FECG
US
US2
The heartbeat coincidence detection feature can be enabled/disabled from a service
setup mode. Refer to the
When heartbeat coincidence detection is enabled, and the monitor detects two
heartbeats that appear to be coinciding, this may indicate that both heart rate
channels are picking up the same signal. When this coincidence occurs, the heart
rate numerics for both heart rates flash alternately (one on while the other is off). As
soon as coincidence is resolved, the numerics stop flashing.
If you disconnect a transducer while coincidence is detected, the numerics stop
flashing.
IMPORTANT
ALARMS PRECEDENCE—Alarms take priority over heartbeat
coincidence. If a fetal heart rate alarm occurs while coincidence is
detected, the pair of numerics stops flashing alternately; only the
alarming parameter numerics continue flashing. If both, fetal heart
rates violate alarm limits, both heart rate numerics flash
synchronously. Fetal heart rate alarms are also indicated by a
flashing
indicator
Alarm
170 Series Monitor
2003023-001
FECG
US
Model 173
Model 173
Model 174
Models 172, 174
"170 Series Service Manual"
and an audio indicator.
US2
Model 174
Models 172, 174
for more information.
5-7