Model 173 Connectors
Ultrasound Connector
FECG Connector
Uterine Activity Connector
3-10
Controls, Indicators, and Connectors: Front Panel Connectors
The ultrasound connector
only a Corometrics ultrasound transducer plug. The fetal heart rate derived from
this transducer shows in the primary fetal heart display.
The FECG connector
a Corometrics FECG cable/legplate plug. The fetal heart rate derived from the spiral
electrode displays in the secondary fetal heart rate display.
The uterine activity connector is a white, round receptacle mechanically keyed to
accept a Corometrics tocotransducer, a Corometrics strain gauge transducer plug, or
any intrauterine pressure catheter with compatible cable plug. The uterine activity
value obtained from this transducer shows in the uterine activity display.
*
If the Model 173 is interfaced to a central information system (CIS), be aware that the CIS may be
designed to alarm when there is no fetal heart rate signal. Therefore it is recommended that you
unplug the ultrasound and/or FECG transducers from the monitor, when not in use, to eliminate false
alarms.
170 Series Monitor
2003023-001
Figure 3-4. Model 173 Connectors
*
is a blue, round receptacle mechanically keyed to accept
*
is a dark grey, round receptacle mechanically keyed to accept