ALARM MESSAGES & EXPLANATIONS (Continued)
CONTROL POWER OFF
The
CONTROL POWER ON
button on the control panel is not engaged, this button should be
illuminated when control power is on.
Figure 33: Control Panel
Press the
CONTROL POWER ON
button & try to clear the alarm. If the alarm does not clear
troubleshoot the control power circuit.
SAFETY CIRCUIT ERROR
The safety circuit has failed allowing control power to be on. The system recognizes this
failure and has alarmed as such. This can be caused by a misplaced jumper or other rewiring
issue. Check the electrical enclosure area for proper wiring of the safety & control circuit.
Figure 34: Electrical Access
40
F-100 Manual Rev H