Table 3.
Integrity
Integrity Indicator Message
Indicator
Number
6
Under Range: Indicates that the input signal is under range. The amplitude of the DUT's signal is
not high enough for the DSP sampler to produce accurate results with the measurement algorithm.
7
Burst Short: Indicates that the burst duration is too short, or part of the burst was not sampled due
to improper triggering.
8
Trigger Early or Fall Early: Indicates that the DUT's burst amplitude fell prematurely or, due to an
early trigger (early relative to a transmitted burst) the measurement sampling operation
terminated before the falling edge of the burst.
9
Trigger Late or Rise Late: Indicates that either the rising edge of the DUT's burst was late or, due
to a late trigger (late relative to a transmitted burst) the measurement sampling operation didn't
start until after the rising edge of the transmitted burst.
10
Signal Too Noisy: Indicates that the measurement algorithm has found the signal measured to be
too noisy to provide accurate results.
11
Sync Not Found: Indicates that the midamble was not found therefore the measurement was not
synchronized.
12
Oven Out of Range: Indicates that a temperature controlled oven (other than the internal timebase
oven) is outside of its operating range. The power meter's oven is checked and its condition
reported with this value. (The internal timebase generates a temporary error message (out of lock)
that is sent to the system error queue and the display. This is not an integrity indicator value, it is
an error message.)
13
Unidentified Error: Indicates errors which are not covered by the other integrity values. Examples
include: parameter errors, algorithm memory errors (too many measurements), measurements
unavailable (unable to control), autorange unable to converge, default calibration data used.
14
PCM Full Scale Warning: Indicates that the PCM signal has reached plus or minus full scale. The
measurement made will be accurate on the PCM signal but would typically indicate an overdriven
or oscillating element in the DUT.
15
Questionable Result for PGSM: Indicates that the user attempted to make an FBER measurement
in a phase 1 system. FBER is only possible in a phase 2 GSM system. This indicator is available
only when the selected broadcast band is PGSM.
16
Questionable Result Due To Channel Mode: Indicates that the channel mode was set to Enhanced
Full Rate Speech while a Decoded Audio measurement was active. Decoded Audio is not supported
for EFR Speech.
S:\Hp8960\E1963A WCDMA\4.0 release\Reference Guide\Chapters\wcdma_prog_gen_integrity.fm
Integrity Indicator
191