Slot Diaphragm; Prerequisites - Siemens SIREMOBIL Compact L Service Instructions Manual

Table of Contents

Advertisement

Adjustments / Programs

Slot diaphragm

NOTE

Prerequisites

The setting of the camera rotation is correctly adjusted and is in the 0° position.
Connect the service PC to the serial interface of the SIREMOBIL.
Start the service program.
Select the "Adjustments, Calibrations..." menu. Answer the query with "Yes".
Select "SCREEN DISPLAY SLOT DIAPHRAGM" in the "Function Groups:" combo box.
Disable circle mask
Switch off zoom.
In the "Available Functions:" window, select "DISABLE CIRCLE MASK" and click on "Ex-
ecute". Close the status window by clicking on the OK button.
For the query "Do you want to reset the Unit", click on "No".
Start adjustment
In the "Available Functions:" window select "INIT" and click on "Execute". Close the sta-
tus window by clicking on the OK button.
For the query "Do you want to reset the Unit", click on "No".
Full format, slot diaphragm open
Move the camera into the 0 degree position.
The X-Iris opens to survey format. To open it further, use the keys for opening the X-Iris.
The slot diaphragm is opened automatically.
Apply fluoro briefly.
Open and close the slot diaphragm display via the keys.
Align the lines shown on the monitor with the front edges of the slot diaphragm blades.
In the window "Available Functions:" select "Slot Diaphragm Open Pos - I.I. Full Format"
and click "Execute".
Close the status window by clicking on the OK button.
Respond to the query "Do you want to reset the unit" by clicking "No". The slot dia-
phragm is closed automatically.
Full format, slot diaphragm closed
Apply fluoro briefly.
Open and close the slot diaphragm display via the keys.
Align the lines shown on the monitor with the front edges of the slot diaphragm blades.
In the window "Available Functions:" select "Slot Diaphragm Close Pos - I.I. Full Format"
and click "Execute".
Close the status window by clicking on the OK button.
Siemens AG
Medical Solutions
Use the lowest KV values possible to ensure optimal detection of
the front edges of the slot diaphragm blades.
SPR2-130.061.02
Rev. 11
Page 17 of 30