● Save alarm.
● Not report pre-alarm.
● Report main alarm with digital output 1.
Procedure
1. Connect the acoustic emission sensor, as described in the "Connecting acoustic emission
2. Press the <PAR> button for five seconds to switch to the parameter menu.
3. Using the
4. Using the
5. Using the
6. Using the
7. Using the
8. Using the
9. Using the
10. Using the
11. Using the
12. Using the
13. Using the
14. Using the
15. Using the
16. Using the
See also
Connecting structure-borne sound sensors (Page 27)
Connecting signal inputs (Page 29)
SITRANS DA400 for material flow monitoring with protection against explosion
Operating Instructions, 09/2007, A5E00841152-01
sensors" section.
and
buttons, set parameter "P. 1".
and
buttons, set the parameter value "on H".
and
buttons, set parameter "P. 5".
and
buttons, set the parameter value "20". A pre-alarm is triggered from
20 dB.
and
buttons, set parameter "P. 6".
and
buttons, set the parameter value "40". A main alarm is triggered from
40 dB.
and
buttons, set parameter "P. 13".
and
buttons, set the parameter value "60". The alarm is triggered after 60
seconds.
and
buttons, set parameter "P. 14".
and
buttons, set the parameter value "St.Ac". Alarms are saved.
and
buttons, set parameter "P. 15".
and
buttons, set the parameter value "oFF". Pre-alarms are not reported.
and
buttons, set parameter "P. 16".
and
buttons, set the parameter value "doAL1". Main alarms are reported
with digital output 1.