Maximum Leveled Power Check Failure; Checking The Output Power - Agilent Technologies E8663B Installation Manual

Analog signal generator
Hide thumbs Also See for E8663B:
Table of Contents

Advertisement

11. Watch the display and ensure that the Unlevel annunciator does not appear at any time during
a full sweep.
The Unlevel annunciator may turn on and off, or remain on; both conditions indicate a failure.
If this occurs, refer to
instructions.
12. Press
.
Sweep > Off
13. Repeat steps 5 through 12 for each frequency range specific to your instrument model number
and options.
Table 3-1 Frequency and Power Level Limits

Maximum Leveled Power Check Failure

If a maximum leveled power check fails, perform the following procedure:
1. Verify the limits for that particular option using the data sheet.
2. If the Unlocked annunciator is displayed, refer to the Troubleshooting chapter of the E8663B
Analog Signal Generator Service Guide.
3. Ensure that the RF OUTPUT connector is connected to a 50Ω load.
4. Ensure that the power level entered corresponds to the value listed in
5. Refer to
Table 3- 2
If the measured power level is more than the power level listed in
knob counterclockwise until the measured power level equals the power level in
6. Note the signal generator front- panel amplitude value and use it to repeat the maximum leveled
power check.
If you have followed the above steps and the check continues to fail, contact the nearest Agilent
Technologies call center for assistance. Refer to
page 10
for contact information.

Checking the Output Power

This test verifies the CW output power from the signal generator. If this test fails, refer to
with Output Power" on page 16
Table 3- 2
shows the recommended equipment for use with the E8663B analog signal generator when
performing this test.
Agilent E8663B Analog Signal Generator Installation Guide
"Maximum Leveled Power Check Failure" on page 13
Amplitude
E8663B
Option 503
15 dBm
Option 509
18 dBm
for the recommended equipment and measure the output power.
for further instructions.
Frequency
Start
Stop
250 kHz
3.2 GHz
3.21 GHz
9 GHz
"Contacting Agilent Sales and Service Offices" on
Operation Verification
Checking the Output Power
for further
# Points
50
100
Table 3-
1.
Table 3-
1, turn the front panel
Table 3-
"Problems
1.
13