A&D UA-767S-W Instruction Manual page 9

Digital blood pressure monitor
Hide thumbs Also See for UA-767S-W:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Symbols
Appears during measurement when
the cuff is attached loosely
Detected rate of IHB/AFib in memory
%IHB/AFib =
Previous measurements stored in
MEMORY.
Average data
FULL BATTERY
The battery power indicator during
measurement.
LOW BATTERY
The battery is low when it blinks.
Unstable blood pressure due to
movement during measurement.
The systolic and diastolic values
are within 10 mmHg of each other.
The pressure value did not increase
during the inflation.
The cuff is not applied correctly.
PUL DISPLAY ERROR
The pulse is not detected correctly.
Blood pressure monitor internal
error
Symbols
Function / Meaning
Number of detected
IHB/AFibs in memory
Total number
English 7
Recommended Action
The reading may yield an
incorrect value. Apply the
cuff correctly, and try the
measurement again.
x100 [%]
Replace all batteries with new
ones when the mark blinks.
Try the measurement again.
Remain very still during
measurement.
Apply the cuff correctly,
and try the measurement
again.
Remove the batteries and
press the
and then install the batteries
again. If the error still
appears, contact the dealer.
button,
START

Advertisement

Table of Contents
loading

Table of Contents