General Specifications

Maximum Voltage Applied to any Terminal ........ 1000 V ac rms or dc
Storage Temperature.......................................... -40 °C to 60 °C (-40 °F to 140 °F)
Operating Temperature ...................................... -20 °C to 55 °C (-4 °F to 131 °F)
Temperature Coefficient ..................................... 0.05 x (specified accuracy) per °C for temperatures <18 °C or >28
Relative Humidity................................................ Noncondensing
Vibration.............................................................. Random, 2 g, 5-500 Hz per MIL-PRF-28800F, Class 2 instrument
Shock .................................................................. 1 meter drop per IEC 61010-1 2
Electromagnetic Compatibility ............................ In an RF field of 3 V/M, accuracy = specified accuracy except in
Safety.................................................................. Complies with ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA-
Certifications ....................................................... CSA per standard CSA/CAN C22.2 No. 61010.1-04; TUV per
Batteries.............................................................. Four AA batteries (NEDA 15A or IEC LR6)
Battery Life.......................................................... Meter use 1000 hours; Insulation test use: Meter can perform at
Size ..................................................................... 5.0 cm H x 10.0 cm W x 20.3 cm L
Weight................................................................. 550 g (1.2 lb.)
IP Rating ............................................................. IP40
Altitude
Operating ........................................................ 2000 m CAT III 1000 V, CAT IV 600 V; 3000 m CAT II 1000 V,
Storage ........................................................... 12,000 m
Over-Range Capability ....................................... 110% of range except for capacitance which is 1 %
Compliance to EN 61557 .................................... IEC61557-1, IEC61557-2
2
°C (< 64 °F or > 82 °F)
0 % to 95 % @ 10 °C to 30 °C (50 °F to 86 °F)
0 % to 75 % @ 30 °C to 40 °C (86 °F to 104 °F)
0 % to 40 % @ 40 °C to 55 °C (104 °F to 131 °F)
sides, oak floor)
temperature: accuracy = specified accuracy ± 5 °C (9 °F). (EN
61326-1:1997).
C22.2 NO. 61010-1-04, and IEC/EN 61010-1 2
measurement category III 1000 V (CAT III) and
CAT IV 600 V.
standard EN 61010
Part 1-1002
least 1000 insulation tests with fresh alkaline batteries at room
temperature. These are standard tests of 1000 V into 1 MΩ with a
duty cycle of 5 seconds on and 25 seconds off.
(1.97 in H x 3.94 in W x 8.00 in L)
CAT III 600 V
nd
Edition (1 meter drop test, six
nd
Edition for