Hach LANGE ORBISPHERE C1100 User Manual page 31

Ozone sensor
Hide thumbs Also See for LANGE ORBISPHERE C1100:
Table of Contents

Advertisement

21. Hold the container steady between thumb
and forefinger of one hand.
22. Lower the sensor into the container until the
top of the anode is covered with electrolyte.
23. Leave for a few seconds to ensure the cotton
washer has fully absorbed some of the
electrolyte and that it is no longer dry.
24. Gently screw the sensor clockwise into the
replacement cartridge, applying minimum
pressure to avoid any damage to the screw
threads.
25. Continue turning until the cartridge is
attached to the sensor, and the sensor is
automatically released from the container.
26. The empty container, the screw top and
packing component can be discarded.
Note: It is normal that some of the electrolyte will
overflow from the replacement cartridge and into the
plastic container.
Maintenance and Troubleshooting
29