Maintenance and Troubleshooting
15. Rinse the sensor head under a tap for 15
16. Do not dry the center electrode area, as the
Important Note: At this stage it is highly
recommended to use the ORBISPHERE 323021
electrochemical cleaning and regeneration center
for improved sensor performance and a
noticeably extended sensor life. Detailed
information on how to use the cleaning and
regeneration center is included in the 32301
Operator Manual. The ORBISPHERE 32301 is a
very efficient cleaning and regeneration tool for
electrochemical sensors. This tool reverses the
electrochemical process that is taking place in the
sensor cell during normal operation. This
removes oxidation and at the same time
regenerates the surface of the electrodes. In
addition, the regeneration center offers a
continuity tester for checking the sensor
electronics.
Note: During this next step, it is very important to
ensure your finger does not come into contact with the
cathode (golden surface) as it could leave greasy
deposits on the surface.
17. Take a new cotton washer from the recharge
18. Place the recharge cartridge container on a
19. Remove the packing component from the
20. If there are any visible bubbles in the
28
seconds, aiming the jet of water directly onto
the sensor head.
gap between cathode and guard should be
left filled with water.
kit. Hold it between thumb and forefinger and
place it on top of the anode.
flat work surface and, keeping the container
upright to avoid spilling any of the electrolyte
inside, carefully unscrew the top.
center of the cartridge, and make sure that
the O-ring remains in place on top of the
cartridge. If it comes away then replace it
before continuing.
electrolyte, remove them using a stirring
motion with the packing component.