Frequency Measurement Test - Fluke ProcessMeter 789 Calibration Manual

Hide thumbs Also See for ProcessMeter 789:
Table of Contents

Advertisement

Frequency Measurement Test

1. Put the calibrator in Standby (STBY) mode.
2. Put the UUT rotary switch in the  (ac volts) position.
3. Press h to toggle to the frequency measurement function.
4. Connect the calibrator to the COM and
Figure 10.
5. Apply the values from the calibrator shown in Table 12 and compare the readings on
the UUT to the acceptable readings shown. Press R to select the voltage range.
UUT
Range
Voltage
199.99 Hz
400 mV
1999.9 Hz
4 V
19.999 kHz
40 V
UUT
Figure 10. Frequency Measurement Test Connections
Table 12. Frequency Measurement Test
Calibrator Voltage
Minimum Acceptable
and Frequency
150 mV @ 100 Hz
1 V @ 1000 Hz
4 V @ 10 kHz
5520A CALIBRATOR
NORMAL
AUX
V, , ,RTD
A, -SENSE, AUX V
HI
LO
GUARD
20A
20V PK MAX
TC
terminals on the UUT as shown in
Reading
Acceptable Reading
99.98 Hz
999.8 Hz
9.998 kHz
5522A
SCOPE
OUT
PREV
STBY
OPR
EARTH
EXGRD
SCOPE
MENU
µ
dBm
sec
7
8
9
SETUP
RESET
m
V
Hz
TRIG
n
W
¡F
NEW
4
5
6
CE
k
A
¡C
REF
p
MEAS
MORE
1
2
3
M
F
TC
MODES
MULT
DIV
/ +
0
SHIFT
ENTER
÷
x
20V PK MAX
ProcessMeter™
Performance Verification
Maximum
100.02 Hz
1000.2 Hz
10.002 kHz
EDIT
FIELD
POWER
I
O
adm010F.EPS
25

Advertisement

Table of Contents
loading

This manual is also suitable for:

Processmeter 787b

Table of Contents