Ac Volts Measurement Test - Fluke ProcessMeter 789 Calibration Manual

Hide thumbs Also See for ProcessMeter 789:
Table of Contents

Advertisement

789/787B
Calibration Manual
Range
4 V dc
4 V dc
40 V dc
40 V dc
400 V dc
400 V dc
1000 V dc
1000 V dc

AC Volts Measurement Test

To prevent possible electrical shock, fire, or personal injury:
1. Put the calibrator in Standby (STBY) mode.
2. Put the UUT rotary switch in the S position.
3. Connect the calibrator to the COM and
4. Apply the values from the calibrator shown in Table 11 and compare the readings on
Range
400 mV ac
400 mV ac
4 V ac
4 V ac
4 V ac
40 V ac
40 V ac
400 V ac
400 V ac
1000 V ac
1000 V ac
24
Calibrator DC Voltage
1 V
3 V
10 V
30 V
100 V
300 V
100 V
800 V
Some of the calibration verification tests involve the use of high
voltages and should be performed by qualified personnel only.
Always place the calibrator in the Standby (STBY) mode between
tests and before handling the test connections or test cables.
Figure 9.
the UUT to the acceptable readings shown.
Calibrator Voltage and Frequency
100 mV @ 60 Hz
300 mV @ 60 Hz
1 V @ 60 Hz
2 V @ 60 Hz
3 V @ 60 Hz
10 V @ 60 Hz
30 V @ 60 Hz
100 V @ 60 Hz
300 V @ 60 Hz
100 V @ 60 Hz
800 V @ 60 Hz
Table 10. DC Volts Test
Minimum Reading
0.998 V
2.996 V
9.98 V
29.96 V
99.8 V
299.6 V
99
798
 Warning
terminals on the UUT as shown in
Table 11. AC Volts Test
Minimum Acceptable
Reading
98.9 mV
297.5 mV
0.991 V
1.984 V
2.977 V
9.91 V
29.77 V
99.1 V
297.7 V
97
792
Maximum Reading
1.002 V
3.004 V
10.02 V
30.04 V
100.2 V
300.4 V
101
802
Maximum Acceptable
Reading
101.1 mV
302.5 mV
1.009 V
2.016 V
3.023 V
10.09 V
30.23 V
100.9 V
302.3 V
103
808