Quick Setup

WARNINGS:
• Installation shall only be performed by qualified personnel and in
accordance with local governing regulations.
• All HART versions of SITRANS LG200 transmitter operate at voltages of
11 to 36 V DC at the device terminals. Higher voltage will damage the
transmitter.
To program the Quick Start parameters, it is important to know the following information
(space is left for recording the values):
1. Probe Model [See Probe Nameplate (7MLxxxx-x) (7xX) for details]: _______________
2. Probe Mounting Type NPT, BSP , or flanged [use part number to establish Probe Mount
type (NPT and BSP differ in insertion length)]: _________________________
3. Measurement Type (level only, volume, interface level, or interface level and volume):
________________________
4. Level Units (inches, centimeters, feet, meters): _________________________
5. Probe Length [listed on model information at top of probe(Note Y01 = xxx cm)]:
____________________
6. Level Offset (desired level reading when liquid is at the end of the probe): _________
7. Dielectric (dielectric constant range of process medium, upper layer dielectric for
interface applications): _________________________
8. Loop Control (Is output current to be controlled by level or volume?): ______________
1
9. Set 4.0 mA
(What is the reference point for the 4.0 mA value?): __________________
1
10. Set 20 mA
(What is the reference point for the 20.0 mA value?): ____________
1.
For more information, see page 65.
Page 42
2
10
1
7
9
8
SITRANS LG200 – INSTRUCTION MANUAL
4
5
6
7ML19985KA02