Configuring The Half-Bridge Completion Jumpers - National Instruments SCXI-1121 Calibration Procedure

Hide thumbs Also See for SCXI-1121:
Table of Contents

Advertisement

1 Thumbscrews
© National Instruments Corporation
3.
Remove the side plate of the SCXI chassis.
4.
Install the SCXI-1121 into the right-most slot of the SCXI chassis.
S C
X I
M A
IN
F R
A M
E
L
N A
T S
T I O
E N
U M
N A
T R
I N S
2
1
2 Right-Most Slot
Figure 2. Side Plate Removal and Module Installation
You do not need to cable the SCXI-1121 to a DAQ device. Leave the
configuration of the digital jumpers W32, W38, and W45 unchanged
because they do not affect this calibration procedure.

Configuring the Half-Bridge Completion Jumpers

Verify that the half-bridge completion network is disabled prior to
calibrating the SCXI-1121. Refer to Figure 3 for the location of the
half-bridge completion jumpers. Refer to Table 2 for the correct jumper
settings to disable the completion network.
3 Side Plate
5
3
SCXI-1121 Calibration Procedure