2-Point Calibration; Setting Alarms - Omega SP-010 User Manual

Load cell smart probe
Table of Contents

Advertisement

4.3.1 2-Point Calibration

A 2-point linearization correction may be applied using either known scaling values or reference weights. 2-Point
Calibration can be performed as described below:
Step 1:
From the Load Cell Input interface on SYNC, click Calibrate.
Step 2:
Place your low reference and click Capture next to Low Reading. This will capture the measurement for
the low reference in the low reading box. Enter your actual low reference weight in the Low Actual Box.
Step 3:
Place your high reference and click Capture next to High Reading. This will capture the measurement for
the high reference in the high reading box. Enter your high reference weight in the High Actual Box.
Step 4:
Click Calibrate to save your calibration.
Note:
Clicking Calibrate or Clear Calibration will clear the TARE offset.
4.4

Setting Alarms

Alarms are set by clicking the
the threshold and alarm type in the Condition section and then select which output to turn on in the Action section. The
alarm can be set to be latching or non-latching in the Recovery section.
icon in SYNC on the desired input signal found in the Inputs Configuration Tab. Setup
10 |
P a g e