Siemens SIPROCESS GA700 Quick Start Manual page 110

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Commissioning: Sequence
8.7 [1.5] Calibration
4. Specify the scope of the calibration: "Single/total" parameter field.
By setting this parameter you define the scope of the calibration - one or all measuring
ranges. The measuring range set in the calibration is always calibrated:
– If you set "Single", calibration is only applied for the set measuring range. Calibrations
– If you set "Total", the calibrations of the other measuring ranges are overwritten by the
In validation, only the set, "leading" measuring range is considered.
5. Assign measuring range(s): "Measuring range" parameter field.
Note
Setting the measuring ranges
Check the settings of the measuring ranges before you assign a range to a calibration.
Make sure that the measuring ranges are set appropriately for the respective application.
Adjust the settings of the measuring ranges again, if necessary: [1.4] Measuring ranges
(Page 103) menu.
If you have set "Total" for the scope in step 4, the selected measuring range acts as the
"leading measuring range".
6. Enter setpoint(s):
Define suitable setpoints for each available component. You must enter one setpoint for
each component with zero gas calibration and span gas calibration. You must enter one
setpoint for zero gas and span gas for each component with 2-point calibrations.
To do so, use the parameter fields "Zero gas setpoint component" and/ or "Span
gas setpoint component".
7. Start the wizard for a calibration or validation.
– "Start validation" parameter field
– "Start calibration" parameter field
Follow the instructions on the display. The success of the calibration/validation is displayed
in the result window once the process is complete.
Rectifying errors
If the device aborts a calibration due to errors, an error code, e.g., "F0123", and a plain text
error message are displayed in the result window.
1. Write down the error code.
2. If you cannot rectify the cause of the error yourself, forward the error code to the service
technician in charge.
110
of the other measuring ranges remain unchanged.
calibration of this "leading" measuring range.
A validation evaluates the current device status with respect to the last calibration. The
need for a new calibration can be derived from the validation results.
Compact Operating Instructions, 05/2015, A5E31805656-03
Quick Start

Advertisement

Table of Contents
loading

Table of Contents