7
General Maintenance
To Install the EI HES
4 Snap the source into place so that the source mount tabs are flush against
5 Connect the ion source wires. (See
6 Install and hand-tighten the source thumbscrews. Do not overtighten the
7 Close the front analyzer chamber. (See
8 Pumpdown the MS. (See
9 Tune the MS. (See
186
the source mount attachment surface of radiator. Some force is required to
overcome the resistance of the electrical contacts, spring loaded surfaces.
EI HES"
on page 161.)
thumbscrews. (See
Figure 60
Chamber"
on page 195.)
"To Autotune the MS for EI Mode"
"To Connect or Disconnect Wiring to the
on page 185.)
"To Close the Front Analyzer
"To Pumpdown the MS"
7000/7010 Series TQ GC/MS Operating Manual
on page 120.)
on page 117.)