Teledyne OLCT 60 User Manual page 57

Fixed pointgas monitor
Table of Contents

Advertisement

Measuring
Gas
range
(ppm)
O3
Ozone
PH3
Phosphine
SiH4
Silane
Sulphur
SO2
Dioxide
CMV
(→30/11/2018)
Vinyl chloride
CMV
(01/12/2019→)
Chloro-
CH3Cl
methane
Dichloro-
CH3Cl
methane
Freon R12
1 % vol.
Freon R22
2000
Freon R123
2000
FX56
2000
Freon R134 a
2000
Freon R134 a
2000
Freon R11
1 % vol.
Freon R23
1 % vol.
Freon R143 a
2000
Freon R404 a
2000
Freon R507
2000
Freon R410 a
1000
Freon R32
1000
NPO60GB
Revision F.2
Temperature
ADF
IS
range
sensor
sensor
(°C)
1.00
0 to +40
1.00
-20 to +40 20 – 90
50.0
-20 to +40 20 – 95
10.0
-20 to +50 15 – 90
30.0
100
200
0 to +40
200
-20 to +40 15 – 90
500
-20 to +55 20 – 95
500
-20 to +55 20 – 95
-20 to +55 20 – 95
-20 to +55 20 – 95
-20 to +55 20 – 95
-20 to +55 20 – 95
-20 to +55 20 – 95
 (IR)
-20 to +50
-20 to +55 20 – 95
-20 to +55 20 – 95
-20 to +55 20 – 95
-20 to +55 20 – 95
-20 to +55 20 – 95
-20 to +55 20 – 95
-20 to +55 20 – 95
FIXED POINTGAS MONITOR
Accuracy (ppm)
% HR
@20°C
+/- 0.03 (from 0 to 0,2
ppm)
10 – 90
+/- 0.05 (from 0.2 to 1
ppm)
+/- 0.05
+/- 1.0
+/- 0.7 (range 0-10)
+/- 5% (from 20 to 70%
15 – 90
FS)
+/- 5% (from 20 to 70%
FS)
+/- 15% (from 20 to 70%
PE)
+/- 15% (from 20 to 70%
PE)
+/- 15% (from 20 to 70%
PE)
+/- 15% (from 20 to 70%
PE)
+/- 15% (from 20 to 70%
PE)
+/- 15% (from 20 to 70%
PE)
+/- 15% (from 20 to 70%
PE)
+/- 40 (from 0 to 50% FS)
0 – 95
+/- 100 (from 50 to 100%
FS)
+/- 15% (from 20 to 70%
FS)
+/- 15% (from 20 to 70%
PE)
+/- 15% (from 20 to 70%
PE)
+/- 15% (from 20 to 70%
PE)
+/- 15% (from 20 to 70%
PE)
+/- 15% (from 20 to 70%
PE)
+/- 15% (from 20 to 70%
PE)
OLCT 60
USER MANUAL
Average
Storage
Resp. Time
LifeExpe
T50/T90
time and
ctancy
(s)
conditions
(months)
18
40/120
(c)
18
30/120
(a)
18
25/120
(a)
36
15/45
(a)
24
10/50
(a)
24
10/50
(a)
40
25/90
(d)
40
25/90
(d)
40
25/90
(d)
40
25/90
(d)
40
25/90
(d)
40
25/90
(d)
40
25/90
(d)
60
40/170
(e)
40
25/90
(d)
40
25/90
(d)
40
25/90
(d)
40
25/90
(d)
40
25/90
(d)
40
25/90
(d)
40
25/90
(d)
51

Advertisement

Table of Contents
loading

This manual is also suitable for:

Olct 60dOlct 60idOldham simptronics olct 60

Table of Contents