OLCT 60
FIXED POINTGAS MONITOR
USER MANUAL
Step
1i.
The "FAULT" and OK LEDs flash one after the other.
1j.
To stop the test or go to next test, Place the magnet over
.
1k.
LCD test display.
To run the test, Place the magnet over
1l.
The display will flashing.
1m.
To stop the test or go to next test, Place the magnet over
.
1n.
Display of the 4-20mA output test.
To run the test, Place the magnet over
1o.
Define the output current value available from the OUT
terminal ( from 1 to 23 mA).
1p.
Setting the tens digit
Set the tens value by placing the magnet over
tens digit flashes. Each magnet action increments the tens
digit. Validate the tens digit by positioning the magnet over
.
1q.
Setting the unit digit
Set the unit value by placing the magnet over
unit digit flashes. Each magnet action increments the unit digit.
Validate the unit digit by positioning the magnet over
1r.
Setting the tenths digit
Set the tenths value by placing the magnet over
tenths digit flashes. Each magnet action increments the tenths
digit. Validate the tenths digit by positioning the magnet over
.
OLCT 60
1s.
generate the define output current
36
Action
.
.
.
Illustration
. The
. The
.
. The
NPO60GB
Revision F.2