Installation.book Page 33 Monday, January 12, 2004 1:13 PM
CAU TI O N

Preparation for Use

WA RN ING
CAU TI O N
Installation Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Altitude: Up to 3000 m (9840 feet).
Humidity: Up to 90% relative humidity to 40
protected from temperature extremes which may cause
condensation.
To ensure adequate cooling, do not obstruct air vents in the
instrument cabinet.
This instrument is designed for use in Installation Category II and
Pollution Degree 2 per IEC 61010-1 and IEC 60664 respectively.
Ventilation Requirements: When installing the instrument in a cabinet,
the convection into and out of the instrument must not be restricted. If
the total power dissipated in the cabinet is greater than 800 watts,
then forced convection must be used.
This product is suitable for indoor use only.
If this instrument is not used as specified, the protection provided by
the equipment could be impaired. This instrument must be used in a
normal condition only (in which all means for protection are intact).
Power Requirements
The instrument requires a power source of 100 to 240 V ac at a
frequency between 47 and 63 Hz (nominal).
Mains supply voltage fluctuations should not exceed +/- 10% of the
nominal line voltage.
Power consumption: 250VA.
Installation (Getting Started)
o
C, but it should be
2
33