Withings BPM Core Instruction Manual page 31

Table of Contents

Advertisement

Available languages

Available languages

Declaration – electromagnetic emissions and immunity
for equipment and systems that are not life-supporting and are specified for use only in a shielded location
The Smart blood pressure monitor, with ECG & digital stethoscope declaration electromagnetic immunity
The Smart blood pressure monitor, with ECG & digital stethoscope system is intended for use in the electromagnetic environment specified below.
The customer or the user of the Smart blood pressure monitor, with ECG & digital stethoscope system should assure that it is used in such an
environment.
Immunity test
Conducted
RFIEC 61000-4-6
Radiated RF IEC 61000-4-3
IEC 60601 test level
Compliance level
3 Vrms
N/A
150 kHz to 80 MHz
N/A
3 V/m
N/A
80 MHz to 2.5 GHz
Electromagnetic environment – guidance
N/A
Portable and mobile RF communications equipment should
be used no closer to any part of the EQUIPMENT or SYSTEM
including cables, than the recommended separation distance
calculated from the equation applicable to the frequency
of the transmitter. Interference may occur in the vicinity
of equipment marked with the following symbol.