Withings BPM Core Instruction Manual page 30

Table of Contents

Advertisement

Available languages

Available languages

RF Statement
Medical Electrical Equipment needs special precautions
regarding EMC and needs to be installed and put into service
according to the EMC information provided in the following.
| Interference may occur in the vicinity of equipment marked
with | Portable and mobile RF communication equipment
(e.g. cell phones) can affect Medical Electrical Equipment. |
The use of accessories and cables other than those specified
may result in increased emissions or decreased immunity
| The device uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment |
The device is suitable for use in all establishments, including
domestic establishments and those directly connected to
the public low voltage power supply network that supplies
buildings used for domestic purposes | Portable and
mobile RF communications equipment should be used no
closer to any part of the device, including cables, than the
recommended separation distance calculated from the
equation applicable to the frequency of the transmitter | The
device is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. Any other
accessories, transducers and cables may result in increased
emissions or decreased immunity and EMC performance.
| The device should not be used adjacent to or stacked
with other equipment and that if adjacent or stacked use
is necessary, which should be observed to verify normal
operation in the configuration in which it will be used.
Medical Electrical Equipment needs special precautions
regarding EMC and needs to be installed and put into service
according to the EMC information provided in the following.
Portable and mobile RF communication equipment (e.g. cell
phones) can affect Medical Electrical Equipment. The use of
accessories and cables other than those specified may result
in increased emissions or decreased immunity of the unit.
Guidance and manufacturer's declaration-electromagnetic emissions
The Smart blood pressure monitor, with ECG & digital stethoscope is intended for use in
the electromagnetic environment specified below. The customer or the user of the Smart
blood pressure monitor, with ECG & digital stethoscope should assure that it is used in
such an environment.
Emissions test
Compliance
CE emissions
Group 1
CISPR11
RE emissions
Class B
CISPR11
Harmonic emissions
Not applicable
IEC 61000-3-2
Voltage fluctuations/
Flicker emissions
Not applicable
IEC 61000-3-3
Electromagnetic
environment - guidance
The Smart blood pressure monitor, with ECG &
digital stethoscope uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
The Smart blood pressure monitor, with ECG
& digital stethoscope is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public
low-voltage power supply network that supplies
buildings used for domesticpurposes.
30

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wpm04

Table of Contents