6.6 Adjusting the blanking value

6.6
Adjusting the blanking value
To change the stored blanking value, go to the 'bL' display (as described in Adjustments
(Page 25)), and proceed as follows:
1. Press the "20" key to increase the blanking value, or the "4" key to decrease it until the
correct value is displayed.
2. Release the key. The display automatically returns to the Run mode (6 secs).
28
3
Indicates that the device is now in blanking mode.
seconds
The stored blanking value is shown
(for example 0.25 m)
Press "20" to increase blanking value
(i.e. 0.36 m)
Press "4" to decrease the blanking value
(i.e. 0.35 m)
6
The new blanking value is shown.