Download Print this page

Fluke FiberInspector Pro OFTM-5352 Instruction Sheet page 3

Video probe

Advertisement

A Use the adapter cable provided to connect the probe to the video input jack
on the side of the tester.
B Screw an adapter tip that matches the connector type being inspected onto
the fiber probe.
C Clean the connector to be inspected.
D Press f . If the tester shows the message "Camera Image Unavailable"
check the connections between the probe and the tester.
E Place the probe on the fiber connector. Turn the larger ring on the probe to
focus the image. Turn the smaller ring to change magnification.
The tester's softkeys provide additional functions. Press
information about using the probe.
F To save the image, press s , select or create a fiber ID; then press s
again.
C
E
F1
F2
SETUP
FUNCTIONS
FIBER
INSPECTOR
D
Figure 1. Using the Probe