Fresenius Medical Care 460017 Operator's Manual page 72

Dry acid dissolution unit
Table of Contents

Advertisement

portable and mobile RF communications equipment and the 99 Gallon Dry Acid Dissolution Unit.
The 99 Gallon Dry Acid Dissolution Unit is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the 99 Gallon Dry Acid Dissolution Unit can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the 99 Gallon
Dry Acid Dissolution Unit as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects, and people.
66
Recommended separation distances between
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2 √P
0.12
0.38
1.2
3.8
12
m
80 MHz to 800 MHz
d = 1.2 √P
0.12
0.38
1.2
3.8
12
800 MHz to 2.5 GHz
d = 2.3 √P
0.23
0.73
2.3
7.3
23
P/N 460017 Rev. D