R I N S E
C Y C L E
5. When the RECIRCULATE Operation is complete, the Dry Acid
Dissolution Unit will switch to DRAIN Operation and empty rinse water
from the Dry Acid Dissolution Unit to the floor drain.
6. At the completion of the DRAIN Operation, the Dry Acid Dissolution
Unit will refill to the 25 Gallon Sensor. The Fill Indicator Light will turn
ON and the second RINSE Operation will start.
7. When the second RINSE CYCLE is finished, the Dry Acid Dissolution Unit
will go to the CYCLE COMPLETE Operation. The Rinse Cycle Complete
Indicator Light will turn ON and the Drain Valve will remain OPEN,
allowing any residual rinse water to go down the drain.
P/N 460017 Rev. D