The PAUSE State function is to allow for special occurrences in the operation of the Dry Acid
Dissolution Unit. For example; if the unit is not working properly or the operator needs to PAUSE any
operation, then the PAUSE button may be pressed.
This will cause the Indicating Light for the current step of
the operation to flash. Pressing the PAUSE button will
disable the pump, drain valve, fill valve, etc. To continue
the cycle, press the START button and the timers will
resume.
The FILL Operation cannot be paused.
CAUTION! D
O NOT ALLOW THE
BACTERIAL GROWTH MAY OCCUR
9.2
S
M
TEP
ODE
CAUTION! STEP MODE
The STEP MODE function is intended to be used during the Disinfection Operation or when it is
necessary to discard an incorrectly mixed batch of solution. The STEP MODE function has the ability to
skip critical steps in the Dissolution and Rinse cycles but using this mode during these cycles can cause
unintended operation of the Dry Acid Dissolution Unit resulting in incorrectly mixed solution.
To enter the STEP MODE, press the STEP MODE ON/OFF
button until the Step Mode Indicator Light illuminates.
The system will enter the STEP MODE and all operations
will be suspended.
Press the STEP button to step to the desired operation.
When the desired operation light becomes illuminated,
press the ON/OFF button and the operation is continued.
CAUTION! W
P/N 460017 Rev. D
O

PERATIONS

U
NIT TO REMAIN FULL OF WATER WITHOUT THE ADDITION OF
.
NOT
FUNCTION IS
TO BE USED TO SPEED UP THE DISSOLUTION OR RINSE
S
L
,
HEN THE
TEP
IGHT IS OFF
THE
D
A
D
U
RY
CID
ISSOLUTION
NIT WILL RESUME OPERATION
D
A
P
.
RY
CID
RODUCT
O
.
PERATIONS
.
27

Advertisement

Table of Contents
loading

Table of Contents