Section 6: Rinse Cycle - Fresenius Medical Care 460017 Operator's Manual

Dry acid dissolution unit
Table of Contents

Advertisement

S
6: R
C
ECTION
INSE
YCLE
CAUTION!
B
EFORE INITIATING THE RINSE CYCLE
P
L
ORT
ID IN PLACE UNLESS MANUAL INSTRUCTS OTHERWISE
CAUTION!
I
F A
WITH TREATED WATER
BE COMPLETED BEFORE THE DISSOLUTION CYCLE IS STARTED
A
P
. D
CID
RODUCT
4
FOR
OR MORE HOURS SINCE THE
1. Before initiating the RINSE CYCLE, the operator must ensure that the Dry Acid Dissolution Unit Access
Port Lid in place (do not fully secure) and the Main Transfer Ball Valve is in the CLOSED position. Make
sure input water supply valve is OPEN ( Figure 1).
2. The RINSE CYCLE consists of two RINSE Operations. Press the Rinse START button. The Fill Indicator
Light will illuminate indicating the RINSE CYCLE has started and the Dry Acid Dissolution Unit is in FILL
Operation (Figure 2). Lift the Access Lid and look into the tank to verify that the tank is filling.
.
3. Once the 25 Gallon Level Sensor is reached, the process automatically
steps to RECIRCULATE Operation (Figure 3). The operation will run for
a twelve (12) minute period. During this operation, the Spray Ball and
jets shall operate in 3 minute intervals. With your safety glasses on,
lift the lid after the Spray Ball has stopped operating and ensure that
the water is level with 25 Gallon Level Sensor. Since the water will be
swirling in the tank it is okay for the level to be a little above the 25
Gallon Sensor, but you should ensure that the level is not rising. During this time period, follow the
procedure outlined on the following page to inspect and clean the Spray Ball Ring. If needed, see
Figure 4: Spray Ball Ring Inspection / Cleaning Instructions.
P/N 460017 Rev. D
,
D
A
D
U
RY
CID
ISSOLUTION
NIT IS FOUND TO BE STANDING FOR
,
THEN A DISINFECTION CYCLE
N
L
D
A
D
O
OT
EAVE THE
RY
CID
ISSOLUTION UNIT WITH ONLY TREATED WATER OR WETTED WITH ONLY TREATED WATER
D
A
D
RY
CID
ISSOLUTION
Figure 2
D
H
THE
RAIN
OSE MUST BE PLACED ABOVE A SUITABLE DRAIN
.
4
(
8.3 S
H
SEE SECTION
ODIUM
YPOCHLORITE
. I
MMEDIATELY AFTER A COMPLETED
U
NIT IS SUSCEPTIBLE TO BACTERIAL GROWTH
Access Port Lid
Figure 1
The Dry Acid
Dissolution
Unit will fill
to the 25
GALLON
LEVEL
SENSOR.
.
. A
LWAYS
OR MORE HOURS WITH TREATED WATER OR WETTED
(
)
BLEACH
DISINFECTION
R
C
,
INSE
YCLE
MAKE A BATCH OF
.
CLOSE Main Transfer Ball
Valve by turning the handle
Counter Clockwise until it
stops.
Spray Ball
25-Gallon
Sensor
Figure 3
L
A
EAVE
CCESS
,
23),
PAGE
MUST
D
RY
11