Changing the Motorized Shutter Operated with the FL SHUTTER Button
4.2.2
You can change the motorized shutter to be operated with the FL SHUTTER button on the Ni-E main body.
Setting Item
[FL SHUTTER
[INTSL]
BUTTON]
[FL TURRET] (default)
[FL 2nd]
[EPI]/[DIA]/[AUX]
[Procedure]
[MIC SETUP] → [BUTTON FUNC]
Choose the shutter to be operated among the options in the
[FL SHUTTER BUTTON].
The displayed buttons depend on the connection
configuration of the motorized shutter.
After making changes to the settings, be sure to press the [SAVE] button on the [MAIN] screen to save the settings to the
microscope's memory. If the microscope is turned off without saving settings, previous values are restored. For more
information on saving, see "1 Bulk Saving of Settings ([MAIN] Screen)" in this chapter.
Chapter 6
Setting Value
Built-in shutter in the motorized HG precentered fiber
illuminator
Built-in shutter of the motorized epi-fluorescence cube
turret
Built-in shutter of the motorized epi-fluorescence cube
turret (2nd layer)
Shutter configured as EPI, DIA, or AUX on the
[CONNECTION] screen
(You must change the connection configuration of the
motorized shutter if the motorized shutter options shown
here are not correct. See "3.2 Configuring the Connection
of Motorized Shutter" in this chapter.)
1-105
Settings
Shutter to Be Operated